17
0.00ppm
UP DWN SET
ESC
1UPIncrease CL
2
value
2DWN Decrease CL
2
value
3 SET The value entered is the CL
2
residual at the time the STR pushbutton was pressed.
4 ESC Go back to the get sample screen
CALIBRATING THE ANALOG OUTPUT
For one to calibrate the analog output, an accurate DMM or equivalent current meter (20mA
range) must be connected alone or in series with the 4 to 20mA (PO1) load.
SET ANALOG OUT
4mA 20mA
EXIT
14mA Set 4mA point
2 20mA Set 20mA point
3 No function
4 EXIT Go back to the calibrate menu.
SET XX.00mA OUT
UP DWN
SET
The XX.00 is either 4.00 or 20.00 and does not change when setting the selected point.
Observe the current reading from the DMM and use the up and down pushbuttons to set the
selected point to 4/20mA.
1UPIncrease current
2DWN Decrease current
3 No function
4 SET Store new 4mA or 20mA point