RA-1000
RESIDUAL ANALYZER
Installation and Operation
MANUAL
Vista de pagina 0
1 2 3 4 5 6 ... 18 19

Indice de contenidos

Pagina 1 - Installation and Operation

RA-1000RESIDUAL ANALYZERInstallation and OperationMANUAL

Pagina 2

10ANALYZER. WHEN THIS ASSEMBLY IS WITHDRAWN, THESESPHERES WILL DROP OUT.1. Refer to OPERATION, Section B and proceed with Steps 1through 7. Remove t

Pagina 3

11E. MOTOR REPLACEMENT1. If the motor is to be replaced, remove the striker from the motorshaft. Install on new motor after turning the center set sc

Pagina 4

12IX. TROUBLESHOOTING CHARTNOTES:A. Verify sample flow at the analyzer intake between the overflow weir andthe filter screen (500 cc/minute) )1 pint/m

Pagina 5

13TROUBLE: Inability to zero.PROBABLE CAUSE: 1. Residual present in sample.CORRECTIVE ACTION: 1. Use de-chlorinated water.TROUBLE: Improper reagent fe

Pagina 6

14MAIN RUN SCREENThe value displayed is the chlorine residual in ppm or mg/l. Below the residual value is a fifty segmentbar-graph representation of t

Pagina 7

15CALIBRATE CELLFrom the residual setup screen, press CELL to enter the cell calibration menu. CALIBRATE CELLZRO SPANESC1 ZRO Zero cell input2 SPAN Se

Pagina 8

16AUTO ZERO SCREENAUTO ZERO 120s.. 0.0mV ESCThis screen will automatically zero the cell input if there is a stable mV input for 120 seconds. If

Pagina 9

17 0.00ppmUP DWN SETESC1UPIncrease CL2 value2DWN Decrease CL2 value3 SET The value entered is the CL2 residual at the time the STR pushbutton was pr

Pagina 10

18SETPOINTSThe RA-1000 has CL2 residual high and low relay (alarm) outputs. Press ALM from theRESIDUAL SETUP screen. Select either the LOW or HIGH set

Pagina 11

19TB5A B1Tx5+ -4-20mA19+ -+ -+ -+ -CELLTHRM+ -PV1+ -PV24-20mA+ -+ -mV12V COMP01SCRPASSRx2SW134motorTB4K1K2K3K42ACPOWER13NGHEAGLE M

Pagina 12

2TABLE OF CONTENTI INTRODUCTION 3II SPECIFICATIONS 3III PRINCIPAL OF OPERATION 4A GENERAL 4B REAGENT FEED SYSTEM 4IV INSTALLATION 5A WALL PANEL MOUNTI

Pagina 13

3I. INTRODUCTIONThe RA-1000 Residual Analyzer/Controller is an amperometric instrumentdesigned to continuously analyze residual levels of free or tota

Pagina 14

4Ambient Temperature 35 degrees to 120 degrees F (2 degrees to 50Range: degrees C)Sample Temperature 35 degrees 120 degrees FRange: (2 degrees to 50

Pagina 15 - 2 residual

5IV. INSTALLATIONA. WALL PANEL MOUNTING1. Position the analyzer panel on a wall at eye level and as close aspossible to the sample source. Secure wit

Pagina 16 - 0.0mV STR

6NOTE: ALL WIRING MUST COMPLY WITH APPLICABLELOCAL AND NATIONAL ELECTRICAL CODES.1. Remove four (4) screws securing side cover and remove cover.2. Co

Pagina 17 - SET XX.00mA OUT

7ANALYZER RANGE 10 grams for 0.5 mg/l 40 grams for 2.0 mg/l 60 grams for 3.0 mg/l100 grams for 5.0, 10, 20 and 30 mg/lc. Shake until crystals are comp

Pagina 18

8E. The analyzer requires a minimum stabilization time of 24 hours. Duringthis time, reagents must be feeding into the cell . Under normalcircumstanc

Pagina 19 - EAGLE MICROSYSTEMS

92. Stop sample flow.3. Lift and rotate clip securing the reagent bottle. Raisereagent bottle approximately 2” and pull down on the valve stemuntil t